The conversion takes place in two steps. First, the complex load impedance must be computed from the complex S-parameter using the characteristic impedance of the measurement, which is 100 Ohms in this case. Next, the S-parameter will be computed from the load impedance using the new reference impedance of 90 Ohms. Because return loss is a function of frequency, this conversion is performed for each frequency.
For example, the complex return loss S11 = 0.53 - 0.12J measured in a 100-Ohm characteristic impedance converts to 90 Ohms:
(see equations)
The relations in Equation 2 were used to convert the insertion- loss measurement in Figure 2 to a 90-Ohm differential impedance. The two curves in Figure 3 show the return loss for 100- and 90-Ohm characteristic impedances.
The differential impedance of mated cable and connectors for the USB3.0 standard can be measured along with the insertion loss using a calibrated TDR system. Calibration is achieved by measuring open, short, and load standards at the reference plane, where the DUT is connected. Using a simple conversion, insertion loss can be used to compensate for the difference in impedance between the measurement system and the DUT.
REFERENCES
1. "Time Domain Spectrum Analyzer and "S" Parameter Vector Network Analyzer," James R. Andrews, Picosecond Pulse Labs application note AN-16a, November 2004.
2. "Converting S-parameters From 50-Ohm to 75-Ohm Impedance," Dallas Semiconductor/Maxim application note, Nov. 21, 2003.